Vital Signs

An alarming result (and that's a very good thing)

By Eliza C. Mackintosh

Staff nurse Kimberly Belanger-Demers checks the monitor.

Amid the bustle of DHMC's orthopaedic unit sits a monitor displaying each patient's vital signs in neat, orderly rows. As a nurse zips by, she hovers for a moment, assesses the colored squares, then bustles off again. If a patient's oxygen level or heart rate should plummet, she knows her beeper will sound and that within seconds she can be at the proper patient's bedside.

"Many people, nationwide, have died because they weren't monitored" continuously, says DHMC's Kenneth Lee. "They could have been rescued." Lee, a clinical manager in biomedical engineering, helped install and test the new monitoring system, Patient SafetyNet, on DHMC's 36-bed orthopaedic unit.

Pagers: The system uses a network of pulse oximeters, which measure heart rate and blood oxygenation through a probe placed on a patient's finger. The devices are linked to the central monitor and to nurses' pagers.

In most hospitals, monitoring in surgical units involves "sampling of intermittent vital signs and clinical examinations," plus closer surveillance of high-risk patients, wrote Drs. Andreas Taenzer and George Blike of DHMC in Anesthesiology. The paper detailed the orthopaedic unit's experience with Patient Safety Net, which is made by a company called Masimo.

Unit: For 11 months before and 10 months after the system's installation, the unit tracked how many patients had to be transferred to the ICU or treated by special resuscitation teams. The authors also compared results on the orthopaedic unit to those of other units.

They found the system significantly reduced rescues and ICU transfers; they estimate it will save 150 ICU bed-days a year. "It's not a given that every safety investment is cost-effective," says Blike, "but in this case it was. The system paid for itself."

The results were so impressive that the project received this year's Health Devices Achievement Award from the ECRI Institute, a nonprofit dedicated to evidence-based patient care. Two additional DHMC surgical units and four medical units have since adopted the system. More are likely to follow, says Jean Avery, senior clinical quality specialist.

They estimate that the system will save 150 ICU bed-days a year.

But the system's success was not a foregone conclusion. On a busy inpatient unit, where one nurse cares for several patients, it is important that alarms sound only when action needs to be taken. Otherwise, nurses can become desensitized to alarms, as the authors of the paper found when they tested the Masimo system on another unit. If it's calibrated for high sensitivity, it can generate too many false alarms—such as if a patient changes position and bumps the pulse oximeter. By the time the nurse gets to the patient's bedside, his or her vital signs would have recovered.

Delays: To combat this problem, the team built in two delays: a 15-second delay before the alarm goes off at the bedside, and an additional 15-seconds before the nurse is paged.

Such monitoring "could potentially become the standard of care," Avery believes, since it reduces costs and complications and keeps ICU beds open for patients who really need them.

She had a chance to experience its benefits from a patient's as well as a nurse's perspective. A relative and a friend had surgery at DHMC during the test period. "I felt better knowing that they had that system there watching them at night," says Avery. "It was comforting."


If you'd like to offer feedback about these articles, we'd welcome getting your comments at DartMed@Dartmouth.edu.

These articles may not be reproduced or reposted without permission. To inquire about permission, contact DartMed@Dartmouth.edu.